Quick answer: Keep two simple records. A receiving log captures each product's name, lot number and expiry date the day it arrives. The client record captures the name and lot number of what you actually used. Together they let you answer a recall in minutes, and they match what Manitoba's nursing college requires and the examples Ontario's nursing college gives.
This is general good practice, not legal advice. Follow your college's standards, the product's instructions for use, and your medical director.
What is a lot number, and why does it matter?
A lot number (Health Canada's term is "control number") identifies a batch made together. The Medical Devices Regulations (MDR) define a control number as "a unique series of letters, numbers or symbols" assigned by the manufacturer, which allows the manufacture, packaging, labelling and distribution of a unit, lot or batch to be traced.
When a manufacturer finds a problem, it recalls by lot. If you know which lot went into which client, you can act quickly and calmly. If you don't, you may have to contact everyone.
What do Canadian regulators and colleges say?
Device labels. Under section 21(1) of the MDR, a device label must show, among other things, "in the case of a Class III or IV device, the control number" and "the expiry date of the device, if the device has one." Lower-class products may carry a lot number too, so check every box.
Distribution records. Section 52(1) requires that "The manufacturer, importer and distributor of a medical device shall each maintain a distribution record." Section 53 explains why: the record "shall contain sufficient information to permit the complete and rapid recall of the medical device." Section 52(2) says this does not apply to "a health care facility in respect of a medical device that is distributed for use within that facility." In other words, the federal distribution-record rule is aimed at suppliers. Your clinic's record comes from good practice and your college.
Nursing colleges. Two recent documents name lot numbers directly:
- The College of Nurses of Ontario (CNO) Aesthetic Services practice guideline (June 2026) lists, under documentation, "recording information about drugs or products used, for example, name, lot number, dose, route and the site used."
- The College of Registered Nurses of Manitoba (CRNM) Practice Direction: Medical Aesthetics (published 10/2025) asks registrants to document lot numbers for "Neuromodulators, dermal fillers, implants and biologics."
Other colleges set general documentation duties. Check your own college's current standard.
Expiry dates. Ontario's Infection Prevention and Control for Clinical Office Practice (Provincial Infectious Diseases Advisory Committee, PIDAC, 1st revision, April 2015) says: "Discard outdated medications. There should be a process in place to check expiry dates before use." For purchased sterile packages with an expiry label, it advises following the "manufacturer's recommendations and discard when outdated."
What should a clinic record, and where?
| Record | When | What to capture | Why |
|---|---|---|---|
| Receiving log | The day a delivery arrives | Date received, product name, manufacturer, catalogue or reference number, lot number, expiry date, quantity, supplier, invoice number, Health Canada licence number (Class II to IV devices) | Links every box on your shelf to its source and its recall status |
| Client record | At each treatment | Product name and lot number, plus what your college requires (for CNO: dose, route and site) | Links each client to a specific lot, as CNO and CRNM describe |
| Expiry check | Weekly, and before each use | Items expiring in the next 60 days, items removed | Meets the PIDAC expectation of a process to check expiry dates |
| Recall check | When an alert arrives | Lot numbers affected, whether you hold them, clients treated | Turns a recall into a search, not a scramble |
How do I set up the receiving log?
Keep it simple enough that it actually gets done.
- One place. A shared spreadsheet or your practice software. One row per lot, not per box.
- Log on arrival. Whoever unpacks the delivery logs it before anything goes on a shelf.
- Shelve by expiry. Put the earliest expiry at the front, so it is used first.
- Flag short dates. Colour any row expiring in the next 60 days.
- Review weekly. Pick a fixed day when you're back in clinic and spend five minutes on the flagged rows.
A template you can copy:
| Date received | Product | Manufacturer | Ref | Lot | Expiry | Qty | Supplier | Invoice | Health Canada licence |
|---|---|---|---|---|---|---|---|---|---|
How do I check for recalls?
The Government of Canada runs a public database of recalls and safety alerts at recalls-rappels.canada.ca, covering health products among others. You can subscribe to email alerts there. When an alert names a device you use, search your receiving log for the lot. If you hold it, set it aside, follow the manufacturer's notice, and talk to your medical director about any affected clients.
If you see a problem with a product, Health Canada's page Report a side effect of a health product, drug or medical device explains how to report it. The CNO guideline also lists "reporting adverse events related to products, drugs and equipment" among the things a practice setting should have processes for.
Where does your supplier fit in?
Your supplier should be able to tell you which lot it shipped you. On Kleos Shop, every device page shows its Health Canada licence number, every invoice lists licence numbers too, and lot numbers can be requested after ordering.
Questions clinics ask
Am I legally required to keep a lot log?
The MDR's distribution-record duty applies to manufacturers, importers and distributors, and section 52(2) exempts health care facilities for devices used in-house. Your college may still expect lot documentation in the client record. This is not legal advice; check your college.
Do Class I and II products have lot numbers?
They may. The MDR requires a control number on the label only for Class III and IV devices, but manufacturers can print one on any class. Log it whenever it is there.
How long should I keep these records?
Follow your college's and province's record-retention rules. They differ by profession and province.
What if a box has no expiry date?
Section 21(1)(g) requires an expiry date only if the device has one. Log "none stated" and follow the manufacturer's instructions for storage and use.
Is a paper log acceptable?
Any format works if it is complete, legible and quick to search. A spreadsheet is easier to search during a recall.
Sources
- Medical Devices Regulations (SOR/98-282), s. 1 ("control number"), laws-lois.justice.gc.ca, and s. 21(1), laws-lois.justice.gc.ca. Justice Laws Website, current to 2026-10-29.
- Medical Devices Regulations (SOR/98-282), ss. 52 to 56 (distribution records). laws-lois.justice.gc.ca
- College of Nurses of Ontario. Aesthetic Services practice guideline. June 2026, p. 8. cno.org
- College of Registered Nurses of Manitoba. Practice Direction: Medical Aesthetics. Published 10/2025, item 29. crnm.mb.ca
- Provincial Infectious Diseases Advisory Committee (PIDAC), Public Health Ontario. Infection Prevention and Control for Clinical Office Practice, 1st revision, April 2015, sections 6.A and 8.I. publichealthontario.ca
- Government of Canada. Recalls and safety alerts. recalls-rappels.canada.ca
- Health Canada. Report a side effect of a health product, drug or medical device. canada.ca