Quick answer: A hypodermic needle has a sharp, bevelled tip. An aesthetic microcannula has a blunt, rounded tip with the opening on the side. Both are sized by gauge, where a higher number means a thinner tube (27G is about 0.40 mm across, 30G about 0.30 mm). A large published study found cannulas were associated with fewer vascular occlusions than needles, but no device removes the risk.
What does needle gauge mean?
Gauge describes the outside diameter (OD) of the tube. The higher the gauge number, the thinner the tube.
Manufacturers publish the diameters. These figures come from Terumo's K-Pack II size table, with BD's technical data sheet as a cross-check.
| Gauge | Nominal outside diameter (Terumo) | BD outside diameter range |
|---|---|---|
| 22G | 0.70 mm | Not listed |
| 23G | 0.60 mm | 0.600 to 0.673 mm |
| 25G | 0.50 mm | 0.500 to 0.530 mm |
| 26G | 0.45 mm | Not listed |
| 27G | 0.40 mm | 0.400 to 0.420 mm |
| 29G | 0.33 mm | Not listed |
| 30G | 0.30 mm | 0.298 to 0.320 mm |
Two International Organization for Standardization (ISO) standards sit behind these numbers. ISO 9626 covers the stainless steel tubing used to make needles, and ISO 6009 sets the hub colour code that lets you identify the diameter at a glance.
Colour is a helpful check, not a guarantee. BD notes that its 16G and 25G PrecisionGlide needles sold in the US "are not compliant to ISO 6009," and Terumo marks one 25G item with a salmon colour on the label rather than the hub. Read the gauge printed on the package.
What does length mean, and how is it written?
Length is the usable length of the needle or cannula, usually printed in millimetres and inches. For example, Terumo lists 27G needles at 12 mm (1/2 in) and 20 mm (3/4 in). Blunt cannulas come longer: TSK lists its STERiGLIDE 22G cannula at 38, 50 and 70 mm.
The right length for a procedure is a clinical choice. Follow your training, the product's instructions for use and your medical director.
How is a needle tip designed?
A hypodermic needle ends in a bevel: the tube is ground at an angle to form a cutting point. Manufacturers refine the grind to lower the force needed to pass through skin. Terumo says its T-Sharp bevel design "decreases penetration resistance." Sol-Millennium describes its insulin syringe needles as "tri-bevel, anti-coring and silicone oil lubricated."
How is a cannula tip designed?
An aesthetic microcannula has a closed, blunt tip. The product leaves through an opening on the side, called the side port.
TSK describes its STERiGLIDE cannula as a "blunt tip micro cannula" with a "dome-shaped tip ... specifically engineered to create less resistance," and with "clear side-port markings." TSK lists these sizes, 20 per box:
| Gauge | Lengths listed by TSK |
|---|---|
| 22G | 38, 50, 70 mm |
| 25G | 38, 50 mm |
| 27G | 25, 38, 50 mm |
| 30G | 25 mm |
How do needles and cannulas compare at a glance?
| Hypodermic needle | Blunt microcannula | |
|---|---|---|
| Tip | Sharp bevel | Closed, rounded |
| Where product exits | End of the tip | Side port |
| Sized by | Gauge and length | Gauge and length |
| Gauges in the ranges cited here | 18G to 30G (Terumo hypodermic range); 27G to 31G (Sol-M insulin syringes) | 22G to 30G (TSK range) |
| Typical lengths | Short (for example 12 mm) | Longer (25 to 70 mm in the TSK range) |
What does published research say about safety?
Two kinds of evidence are often cited. Neither is a technique guide.
A large survey of real-world injections. In a 2021 retrospective cohort study in JAMA Dermatology, Alam and colleagues collected reports from 370 board-certified US dermatologists covering about 1.7 million filler syringes injected between August 2018 and August 2019. They observed 1 occlusion per 6,410 1-mL syringe injections with needles and 1 per 40,882 with microcannulas. In multivariate analysis, "injections with cannula had 77.1% lower odds of occlusion." The authors concluded that both methods were associated with a very low risk, and that cannulas were less often associated with occlusion events. The data were self-reported and retrospective.
Cadaver studies of how much force it takes to enter an artery. Results here are not fully consistent:
- Pavicic and colleagues (Plastic and Reconstructive Surgery, 2019) found that 22G and 25G cannulas needed more force than matching needles to penetrate an artery, while 27G cannulas needed similar force to 27G needles.
- Siperstein and colleagues (Aesthetic Surgery Journal, 2023) tested 22G to 30G in four cadavers and concluded that "all gauge cannulas (including 27 and 30 gauge), require more force than the same size needle to penetrate an artery." They suggested that friction and flexibility, rather than penetration force alone, help keep an instrument outside the vessel.
The fair summary: cannulas were associated with fewer occlusions in a large survey, and the behaviour of small-gauge cannulas is still being studied. Choice of device, technique and scope of practice are governed by your regulatory college, your training, the product's instructions for use and your medical director.
Questions clinics ask
Is a 27G needle thinner than a 25G needle?
Yes. Higher gauge means thinner. 27G is about 0.40 mm across and 25G about 0.50 mm.
Are cannulas risk-free?
No. The 2021 JAMA Dermatology study recorded occlusions with both needles and cannulas, at very low rates for both.
Why is my 25G needle not orange?
Hub colours follow ISO 6009, but not every product complies. BD says some of its US 25G needles do not, and Terumo marks one 25G item with a salmon label colour. Always read the printed gauge.
Do cannulas come in the same gauges as needles?
Many do. TSK's STERiGLIDE range runs from 22G to 30G.
Where does the product come out of a cannula?
From a side port, not from the tip itself.
Does a thinner cannula behave like a needle?
The evidence is mixed. One cadaver study found 27G cannulas needed similar force to 27G needles to enter an artery. A later one found all cannula sizes tested, including 27G and 30G, needed more force than same-size needles.
Sources
- Terumo Europe. K-Pack II needles, size table. terumo-europe.com
- BD. BD disposable hypodermic needle technical data sheet. bd.com
- ISO 6009:2016 Hypodermic needles for single use, colour coding for identification (preview). webstore.ansi.org
- Sol-Millennium Medical. Sol-M Standard Insulin Syringe sell sheet NACA-DC-463-0425.
- TSK Laboratory. STERiGLIDE cannula. shopcan.tsklab.com
- Alam M, Kakar R, Dover JS, et al. Rates of vascular occlusion associated with using needles vs cannulas for filler injection. JAMA Dermatol. 2021;157(2):174-180. doi:10.1001/jamadermatol.2020.5102
- Healio. Vascular occlusion may be less likely with fillers via cannulas vs needles (January 5, 2021). healio.com
- Pavicic T, Webb KL, Frank K, Gotkin RH, Tamura B, Cotofana S. Arterial wall penetration forces in needles versus cannulas. Plast Reconstr Surg. 2019;143(3):504e-512e. doi:10.1097/PRS.0000000000005321. Summary: aestheticmed.co.uk
- Siperstein RD, Broadfield DC, Hansen PY, Conger JR, Zhang-Nunes SX. Arterial wall and tissue penetration force with various cannulas and needles. Aesthet Surg J. 2023;43(12):1521-1529. doi:10.1093/asj/sjad149
- IMCAS. Small gauge cannulas are safe (January 12, 2024). imcas.com